服务热线:400 086 6238-1

新闻动态

热电偶允差表

添加时间:2011/5/9 17:02:30

标准热电偶允差表

 

分度号

类型

Temperature Range ℃
for Standard Limits of Error

误差标准限值的温度范围℃

Standard Limits

Of Error

误差标准限值

Temperature Range ℃

for Special Limits of Error

误差特殊限值的温度范围℃

Special Limits
of Error

误差特殊限值

J

0-293

±2.2℃

0 to 275

± 1.1℃

293 to 760

± 0.75%

275 to 760

± 0.4%

K

-200 to -110

± 2% *


*

-110 to 0

± 2.2℃*


*

0 to 293

± 2.2℃

0 to 275

± 1.1℃

293 to 1250

± 0.75%

275 to 1250

± 0.4%

T

-200 to -67

± 1.5%*


*

-67 to 0

± 1℃*


*

0 to 133

± 1℃

0 to 125

± 0.5℃

133 to 350

± 0.75%

125 to 350

± 0.4%

E

-200 to -170

± 1%*


*

-170 to 0

± 1.7℃*


*

0 to 340

± 1.7℃

0 to 250

± 0.1℃

340 to 900

± 0.5%

250 to 900

± 0.4%

R

0 to 600

± 1.5℃

0 to 600

± 0.6℃

600 to 1450

± 0.25%

600 to 1450

± 0.1%

S

0 to 600

± 1.5℃

0 to 600

± 0.6℃

600 to 1450

± 0.25%

600 to 1450

± 0.1%

B

800 to 1700

± 0.5%

800 to 1700

-

* Thermocouples and thermocouple materials are supplied to meet the limits of error specified or temperatures above 0℃. A thermocouple material may not confirm to the published sub-zero limits of error for that material when purchased, unless conformance is agreed to between customer and supplier.

** Special limits of error for sub-zero temperatures have not yet been established. The following limits for calibrations E and T are useful to start discussion between customer and supplier.

-200 to 0℃
Type E ± 1℃or ± 0.5%, whichever is greater
Type T ± 0.5℃ or ± 0.8%, whichever is greater

Sub-zero limits of error for Type J and sub-zero special limits of error for Type K are not considered because of the characteristics of the materials.

 

ANSI Limits of Error Extension Wire and Compensating Extension补偿导线允差表

(Reference Junction 0℃)

Type

Temperature Range ℃

Standard

Special

JX

0 to 200

± 2.2℃

± 1.1℃

KX

0 to 200

± 2.2℃

-

TX

-60 to +100

± 1.0℃

± 0.5℃

EX

0 to 200

± 1.7℃

-

 

Compensating Extension Wire

 

RX - SX
(Copper/#11 Alloy)

0 to 200

± 5℃†
± 0.057 mV

-

BX
(Copper/Copper)

0 to 100

+0, -3.7℃
+0, - .033mV

-

† The degree C limits of error are based on a measuring junction temperature greater than 870℃ for Types SX and RX, 1000oC for Type BX.

 

Non ANSI Limits of Error Thermocouples非ANSI热电偶允差

 

Type

Temperature Range, ℃

Limits of Error

W-W, 26% Re

0 to 427

± 4.4℃

427 to 2316

± 1%

W, 5% Re-W, 26% Re

0 to 427

± 4.4℃

427 to 2316

± 1%

W, 3% Re-W, 25% Re

0 to 427

± 4.4℃

427 to 2316

±1%

Platinel II
(Platinel 5355/Platinel 7674)

0 to 600

± 0.10mV

600 to 700

± 0.15mV

700 to 1300

± 0.20mV

 

Non ANSI Limits of Error Compensating Extension Wire非ANSI热电偶补偿导线允差

 

Type

Temperature Range, ℃

Limits of Error

Alloy 200/Alloy 226
for W-W, 26% R

0 to 260

± 0.14mV

Ally 203/Alloy 225
for W, 3% Re-W 25% Re

0 to 260

± 0.11mV

Ally 405/Alloy 426
for W, 5% Re-W, 26% Re

0 to 871

± 0.11mV

Chromel P/Alloy 30*
for Platinel 5355/ Platinel 7674

0 to 100

± 1℃

100 to 200

± 6.4℃

*For Extension wire/thermocouple wire Junction temperature (both equal) above about 800 Type K wire may be used with minimal error.

Limits of error apply only to thermocouples supplied by the manufacturer. The calibration of a thermocouple may change because of the influence of such factors as temperature, temperature cycling, and time.